Table 1.

Comparison of measured linear attenuation coefficients μ, and refractive index decrements δ with literature values for the material phantom at an effective energy of Eeff = 24.6 keV for the conventional attenuation image and Eeff = 23.8 keV for the phase-contrast image.

μm [cm−1] μ1 [cm−1] δm [10−7] δ1 [10−7]
NaI (5.9) mg/ml 0.597 ± 0.006 0.609 4.08 ± 0.07 4.07
Blood 0.568 ± 0.006 0.574 4.31 ± 0.07 4.27
Ethanol 0.323 ± 0.006 0.325 3.23 ± 0.06 3.27
PMMA 0.470 ± 0.006 0.470 4.70 ± 0.06 4.70
POM 0.628 ± 0.006 0.628 5.55 ± 0.06 5.55
Nylon 0.423 ± 0.006 0.419 4.61 ± 0.06 4.58
Water 0.523 ± 0.006 0.523 4.07 ± 0.06 4.07

The subscripts m and l indicate measured and literature values. The literature values are calculated from tabulated values from the NIST database25, from26 for blood and from20 for the NaI solution. The given uncertainty includes the standard deviation of the image region and the systematic error.