Amplatzer PFO Occluder Sizing Recommendations

Shortest Distance From Defect to Aortic Root, or Distance From Defect to Superior Vena Cava Orifice Suggested Amplatzer PFO Occluder Size
≥17.5 mm 35 mm
12.5 to 17.4 mm 25 mm
9 to 12.4 mm 18 mm
<9 mm Do not implant device