Table 2.

Precision and accuracy of CAMD assay in rat plasma (n = 6).

Concentration
(ng/mL)
Intra-day Inter-day
Precision
(%, RSD)
Accuracy
(%, RE)
Precision
(%, RSD)
Accuracy
(%, RE)
0.64 7.10 4.30 4.80 3.30
1.80 6.20 -2.30 6.30 4.40
30.00 5.60 5.30 2.90 3.70
49.50 6.90 3.80 4.90 4.90