Skip to main content
. 2018 Oct 1;9:111–115. doi: 10.1016/j.reth.2018.08.003

Fig. 1.

Fig. 1

The construction of experimental chamber in clean room and H2O2 concentrations. (a) Schematic illustration of the experimental environments. An experimental chamber with vinyl sheet was set in a biological clean room including a BSC and H2O2 generator. Filled squares indicate H2O2 sensors, and the dotted line indicates the vinyl sheet. (b) Time profile of H2O2 concentration. Curved line shows the concentration of H2O2 in the vinyl sheet measured by the sensor A in Fig. 1 during the decontamination. (c) Line shows the concentration of H2O2 in the BSC measured by sensor B and c in Fig. 1 during and after decontamination. The minimum detection limit of the sensors was 3 mg/L. Time A: t = 0 is defined as the time point at which H2O2 concentration reached 1 mg/L, as measured by a detection tube. The dotted line indicates the end of 5 h of decontamination. Each arrow indicates the time when the respective dish was uncovered in the subsequent experiments.