Precision and accuracy of analytes at LLOQ and QC samples in beagle dog plasma (n = 6).

Analytes Spiked Concentration (ng/mL) Mean ± SD (ng/mL) Intra-day Precision RSD (%) Inter-day Precision RSD (%) Accuracy RE (%)
aesculin 0.50 0.56 ± 0.15 10.7 6.9 16.5
1.52 1.54 ± 0.22 4.1 8.4 1.6
12.18 12.42 ± 2.16 9.5 6.2 2.8
78.00 74.11 ± 6.30 5.9 4.6 6.8
aesculetin 1.09 1.12 ± 0.15 11.9 14.2 5.1
3.27 14.02 ± 1.57 6.0 7.4 2.4
26.16 24.31 ± 2.23 8.9 9.1 −4.9
167.3 172.3 ± 18.35 3.5 8.9 5.4
fraxetin 2.15 2.08 ± 0.22 15.2 7.7 −1.8
6.45 6.34 ± 0.36 4.5 3.3 −1.5
51.60 53.53 ± 2.31 6.4 7.2 3.7
330.2 347.9 ± 18.46 9.3 7.9 5.4
fraxin 1.25 1.20 ± 0.11 10.5 16.1 −3.4
3.76 3.44 ± 0.36 5.8 3.9 −5.2
30.12 32.90 ± 2.74 7.0 5.6 5.9
192.8 186.3 ± 11.06 5.8 7.3 3.3
polydatin 0.50 1.23 ± 0.13 6.2 15.6 13.1
1.52 1.55 ± 0.39 9.5 11.6 8.3
12.18 13.87 ± 2.67 6.9 5.8 6.4
78.00 75.1 ± 13.39 4.3 7.7 4.8