TABLE 2.
Biograph mCT Spatial Resolution Measured Using Cylinder Technique
| FWHM (mm) |
||
| Protocol | Radial | Axial |
| A | 7.07 ± 0.10 | 7.82 ± 0.12 |
| B | 5.53 ± 0.19 | 5.78 ± 0.09 |
Data are as mean ± SD.
Biograph mCT Spatial Resolution Measured Using Cylinder Technique
| FWHM (mm) |
||
| Protocol | Radial | Axial |
| A | 7.07 ± 0.10 | 7.82 ± 0.12 |
| B | 5.53 ± 0.19 | 5.78 ± 0.09 |
Data are as mean ± SD.