SAXS data collection and analysis parameters

SAS data collection parameters
    Instrument BioCAT facility at the Advanced Photon Source beamline 18ID with Pilatus3 1M (Dectris) detector
    Wavelength (Å) 1.033
    Beam size (μm2) 160 (h) × 75 (v)
    Camera length (m) 3.5
    q measurement range (Å−1) 0.00365–0.383
    Absolute scaling method NAa
    Basis for normalization to constant counts To incident intensity, by ion chamber counter
    Method for monitoring radiation damage Automated frame-by-frame comparison of relevant regions
    Exposure time, number of exposures 1-s exposure time with a 2-s total exposure period of entire SEC elution
    Sample configuration SEC-SAXS. Size separation by an ÄKTA Pure with a Superdex 200 Increase 10/300 GL column. SAXS data measured in a 1.5-mm inner diameter quartz capillary.
    Sample temperature (°C) 20

Software employed for SAS data reduction, analysis, and interpretation
    SAXS data reduction Radial averaging; frame comparison, averaging, and subtraction done using BioXTAS RAW 1.4.0 (55) and ATSAS (56)
Basic analysis: Guinier, molecular weight, P(r) Guinier fit and molecular weight using BioXTAS RAW 1.4.0 (55), P(r) function using GNOM (65)

a Not applicable.