Skip to main content
. Author manuscript; available in PMC: 2018 Nov 12.
Published in final edited form as: J Magn Reson Imaging. 2015 Jun 26;43(1):128–137. doi: 10.1002/jmri.24979

Table 3.

Inter-study reproducibility (comparison of Exam A and Exam B) for global torsion and torsion rates as determined by three different definitions.

Mean difference ± 2SD CoV[%] ICC (95% CI)

Twist Peak Torsion [°] 0.0 ± 3.9 20.3 0.91 (0.74-0.97)
Peak Systolic Torsion Rate [° s-1] 0.7 ± 33.1 26.9 0.84 (0.53-0.94)
Peak Diastolic Torsion Rate [° s-1] 1.7 ± 53.9 43.1 0.36 (0.00-0.78)
Normalized Twist Peak Torsion [° cm-1] 0.0 ± 0.9 19.9 0.92 (0.76-0.97)
Peak Systolic Torsion Rate [° cm-1 s-1] 0.0 ± 7.7 25.9 0.84 (0.54-0.95)
Peak Diastolic Torsion Rate [° cm-1 s-1] 0.3 ± 12.3 40.8 0.47 (0.00-0.82)
CL Shear Angle Peak Torsion [°] 0.0 ± 1.7 19.0 0.90 (0.72-0.97)
Peak Systolic Torsion Rate [° s-1] 0.3 ± 18.7 29.0 0.82 (0.48-0.94)
Peak Diastolic Torsion Rate [° s-1] 0.1 ± 27.9 45.5 0.34 (0.00-0.78)

SD, standard deviation; CoV, coefficient of variation; ICC, intraclass correlation coefficient; CI, confidence interval; CL Shear Angle, circumferential-longitudinal shear angle