. 2018 Sep 5;6(9):e10706. doi: 10.2196/10706

Table 2.

Two-way mixed-model intraclass correlation with measure of consistency.

Walking speed (km/h) n PiezoRx Actigraph Omron


Intraclass correlation (95% CI)a P value Intraclass correlation (95% CI)a P value Intraclass correlation (95% CI)a P value
2.4 43 .71 (0.53 to 0.83) <.001 .16 (−0.15 to 0.44) .88 .14 (−0.16 to 0.42) 0.92
3.2 43 .80 (0.67 to 0.89) <.001 .22 (−0.09 to 0.49) .36 .28 (−0.02 to 0.53)b .04
4.0 42 .98 (0.97 to 0.99) <.001 .70 (0.50 to 0.83) <.001 .94 (0.90 to 0.97) <.001
5.6 41 .99 (0.976 to 0.993) <.001 .96 (0.93 to 0.98) <.001 .39 (0.10 to 0.62)b .03
6.4 38 .95 (0.91 to 0.98) <.001 .93 (0.86 to 0.96) <.001 .53 (0.26 to 0.72)b .01
7.2 30 .99 (0.97 to 0.992) <.001 .98 (0.95 to 0.99) <.001 .66 (0.43 to 0.81)b <.001
Overall N/Ac .97 (0.965 to 0.978) <.001 .72 (0.65 to 0.77) <.001 .62 (0.54 to 0.69)b <.001

aThree decimal places were used for clarity when necessary.

bn=42.

cN/A: not applicable.