Table 2.
The sensitivity and specificity of Preschool Audiometry Screening System speech audiometry app version 1 compared with that of Conventional Play Audiometry.
| Test performance | Conventional play audiometry | |
|
|
Speech reception threshold 20 decibel, % (95% CI) | Pure-tone average0.5-2 KHz 20 decibel, % (95% CI) |
| Sensitivity | 62.50 (28.95-96.05) | 60.00 (29.64-90.36) |
| Specificity | 93.75 (89.27-98.23) | 94.60 (90.30-98.79) |
| Positive likelihood ratio | 10 (4.08-24.49) | 11 (4.35-27.83) |