Table 2.

Alarm sequence.

Serial no. Default setting of the alarm (as released to the hospital floor); total number of alarms=18 Modified to patient condition using Smart SatSec; total number of alarms=11

Alarm type Intervention type Alarm type Intervention type
1 Advisory Nonactionable Removeda Removeda
2 Warning Actionable Warning Actionable
3 System message Nonactionable Removeda Removeda
4 Actionable Actionable Actionable Actionable
5 Warning Actionable Warning Actionable
6 System message Nonactionable Removeda Removeda
7 Warning Actionable Warning Actionable
8 Actionable Actionable Actionable Actionable
9 Warning Actionable Warning Actionable
10 System message Nonactionable System message Nonactionable
11 System message Nonactionable Removeda Removeda
12 Advisory Nonactionable Removeda Removeda
13 Warning Actionable Warning Actionable
14 Advisory Nonactionable Advisory Nonactionable
15 Actionable Actionable Actionable Actionable
16 System message Nonactionable System message Nonactionable
17 Advisory Nonactionable Removeda Removeda
18 Advisory Nonactionable Removeda Removeda

aThese alarms were not presented. Removed alarms: 5 premature ventricular contraction, 1 missed beat, and 1 noninvasive blood pressure.