TABLE VI.
The STD and bias in the Layered Phantom with Uniform Backscatter experiment. In each layer, the smallest values are highlighted in bold font.
Layer 1 |
Layer 2 |
Layer 3 |
||||
---|---|---|---|---|---|---|
LSq | DP | LSq | DP | LSq | DP | |
STD | ||||||
α (dB/cm-MHz) | 0.059 | 0.001 | 0.038 | 0.014 | 0.069 | 0.035 |
b (1/cm-sr-MHzn) | 8.423e-07 | 7.132e-09 | 8.979e-07 | 1.697e-08 | 5.774e-06 | 1.471e-07 |
n | 0.323 | 0.021 | 0.400 | 0.014 | 0.842 | 0.081 |
Bias | ||||||
α (dB/cm-MHz) | 0.022 | 0.009 | 0.028 | 0.004 | 0.091 | 0.028 |
b (1/cm-sr-MHzn) | 1.272e-06 | 3.441e-07 | 4.155e-07 | 1.242e-06 | 5.527e-06 | 5.036e-07 |
n | 0.117 | 0.062 | 0.188 | 0.320 | 0.290 | 0.0139 |