Table 2. Safety and procedural aspects of CrossCountryTM Transbronchial Access Tool.

Site Total number (%) Size, cm (mean ± SD, range) Visibility on fluoroscopy (%) Adverse events Tunnel length of 7 lesions accessed by TBAT, cm (mean ± SD, range)
RUL 4 (18.2) 2.38±1.37, 0.7–3.5 3 (75.0) 0 1.33±0.29, 1–1.5
LUL 11 (50.0) 1.98±0.43, 1.2–2.8 4 (46.4) 0 1.55±0.44, 1–2
RLL 2 (9.1) 3.25±2.76, 1.3–5.2 1 (50.0) 0 1.25±0.35, 1–1.5
LLL 3 (13.6) 1.4±0.26, 1.1–1.6 2 (66.7) 0 1.75±0.35, 1.5–2.0
RML 2 (9.1) 2.05±.5, 1.7–2.4 2 (100.0) 0 1.5±0.71, 1–2

RUL, right upper lobe; LUL, left upper lobe; RLL, right lower lobe; LLL, left lower lobe; RML, right middle lobe; TBAT, Transbronchial Access Tool.