Table 2.
Error detection sensitivity for AC, PDIP and combined processes with three settings of Gamma index for head and neck (H&N) treatment plans. δTotal, δColli, δDose, and δMLC correspond to the detection sensitivity of total error, collimator angulation error, dose error, and MLC error
Protocol | Process | δTotal(%) | δColli(%) | δDose(%) | δMLC(%) |
---|---|---|---|---|---|
3%/3 mm/Global | AC | 54 | 44 | 30 | 87 |
PDIP | 40 | 48 | 10 | 47 | |
Combined | 58 | 52 | 30 | 87 | |
3%/3 mm/Local | AC | 88 | 80 | 90 | 100 |
PDIP | 74 | 60 | 90 | 87 | |
Combined | 92 | 84 | 100 | 100 | |
2%/2 mm/Global | AC | 76 | 68 | 70 | 93 |
PDIP | 50 | 52 | 30 | 60 | |
Combined | 80 | 72 | 80 | 93 |