Error detection sensitivity for AC, PDIP and combined processes with three settings of Gamma index for head and neck (H&N) treatment plans. δTotal, δColli, δDose, and δMLC correspond to the detection sensitivity of total error, collimator angulation error, dose error, and MLC error

Protocol Process δTotal(%) δColli(%) δDose(%) δMLC(%)
3%/3 mm/Global AC 54 44 30 87
PDIP 40 48 10 47
Combined 58 52 30 87
3%/3 mm/Local AC 88 80 90 100
PDIP 74 60 90 87
Combined 92 84 100 100
2%/2 mm/Global AC 76 68 70 93
PDIP 50 52 30 60