Skip to main content
. 2018 Oct 17;19(6):140–148. doi: 10.1002/acm2.12459

Table 2.

Dose volume constraints for organs at risk

Structure Dose limits (Gy[RBE])
Esophagus D33% <65; D67% <55, Dwhole volume ≤45, as low as reasonably achievable
Liver Dwhole volume ≤25; D50% ≤35
Total normal lung V20 Gy[RBE] <37% is desirable; V20 Gy[RBE] >41% is a major deviation
Spinal cord Dmax ≤50
Heart D33% ≤60; D67% ≤45; Dwhole volume ≤30, as low as reasonably achievable; V50 <25%; Dmean <20
Skin Dmax ≤55 (decided by the treating physician)