Skip to main content
. 2018 Oct 17;9(11):5503–5510. doi: 10.1364/BOE.9.005503

Table 3. Phantoms used for SFDI probe evaluation.

Phantom # Expected μa
@ 810 nm (cm−1)
Average μa error for all λs (%) Expected μ's
@ 810 nm (cm−1)
Average μ's error for all λs (%)
1 (ref) 0.0745 0.19 4.9056 0.22
2 0.1293 2.40 4.9056 2.82
3 0.2387 3.87 4.9056 0.72
4 0.0745 9.82 6.8678 9.77
5 0.0745 13.66 7.5546 8.08
6 0.0745 10.52 9.8112 6.43