Published in final edited form as: J Heart Lung Transplant. 2018 Aug 11;37(12):1467–1474. doi: 10.1016/j.healun.2018.08.007

Figure 6. Average temperature changes at four different locations between (A) short- and (B) long- range configuration.

Figure 6.

Thermocouples were placed at four different locations, inside the skin above the receive coil, directly above the receive coil, directly below the receive coil, and core temperature. The greatest temperature rise occurred at directly above the receive coil. All temperature measurements never exceeded the maximum allowable temperature deviation of the body, 2 °C, for an implanted device inside the trunk.