. 2018 Aug 3;39(8):1635–1641. doi: 10.1007/s00246-018-1943-3

Table 2.

Technical characteristics

Low dose
N = 100
Normal dose [14]
N = 17
P value
Exposure parameters
 Tube voltage 3DRA (kV) 70 (60–96) 90 (90–90) < 0.001
 Tube current 3DRA (mA) 228 (53–395) 69 (26–363) < 0.001
 Exposure time 3DRA (ms) 465.5 (333.2–1330) 843.2 (452.2–868) < 0.001
 Skin dose 3DRA (mGy) 9 (1.7–83) 20,86 (10,33–90,69) < 0.001
 DAP 3DRA (mGy·cm2) 1279 (150–16,987) 3128 (1231–17,273) < 0.001
 Procedural time (min) 153 (30–360) 165 (60–540) 0.360
 Total fluoroscopy time (min) 28 (0.4–121) 22 (6.4–81.4) 0.497
Effective dose
 3DRA (mSv) 0.54 (0.12–2.23) 1.62 (0.70–4.94) < 0.001
 Fluoroscopy (mSv) 1.53 (0.00–25.40) 4.4 (0.2–15.8) 0.002
 Angiography (mSv) 0.45 (0.00–19.33) 3.6 (0–79.1) 0.009
 Total catheterization (mSv) 2.64 (0.27–28.13) 12.4 (2–99.9) < 0.001
 Interventional group 3.65 (0.49–28.13)
 Diagnostic group 0.86 (0.27–10.09)

Values represent median and range