. 2018 Oct 10;41(12):2634–2636. doi: 10.2337/dc18-1249

Table 1.

Outcomes of Control-IQ use

Glycemic outcomes of Control-IQ use
Metric Overall Daytime Nighttime
Mean glucose, mg/dL 129 135 121
Coefficient of variation, % (median) 27 27 21
% time <54 mg/dL (median) 0.7 0.0 0.0
% time <60 mg/dL (median) 1.1 2.0 0.0
% time <70 mg/dL (median) 2.9 4.1 1.0
% time in range 70–180 mg/dL (mean) 87 82 94
% time >180 mg/dL (median) 4.7 7.5 5.7
% time >250 mg/dL (median) 0.0 0.0 0.0
% time >300 mg/dL (median)
0.0
0.0
0.0
Participants’ impression of Control-IQ use
Question (1 = lowest rating to 5 = highest rating)
Average score
How easy to use was the device? (1 = Not at all to 5 = Extremely) 5.0
How useful in managing your diabetes was the device? (1 = Not at all to 5 = Extremely) 4.8
How much did you trust the device? (1 = Not at all to 5 = Extremely) 5.0
I had greater peace of mind while wearing the device. (1 = Strongly disagree to 5 = Strongly agree) 4.8
It will be hard to give up the device once the study is over. (1 = Strongly disagree to 5 = Strongly agree) 5.0