Table 1.
Outcomes of Control-IQ use
Glycemic outcomes of Control-IQ use | |||
---|---|---|---|
Metric | Overall | Daytime | Nighttime |
Mean glucose, mg/dL | 129 | 135 | 121 |
Coefficient of variation, % (median) | 27 | 27 | 21 |
% time <54 mg/dL (median) | 0.7 | 0.0 | 0.0 |
% time <60 mg/dL (median) | 1.1 | 2.0 | 0.0 |
% time <70 mg/dL (median) | 2.9 | 4.1 | 1.0 |
% time in range 70–180 mg/dL (mean) | 87 | 82 | 94 |
% time >180 mg/dL (median) | 4.7 | 7.5 | 5.7 |
% time >250 mg/dL (median) | 0.0 | 0.0 | 0.0 |
% time >300 mg/dL (median) |
0.0 |
0.0 |
0.0 |
Participants’ impression of Control-IQ use | |||
Question (1 = lowest rating to 5 = highest rating) |
Average score |
||
How easy to use was the device? (1 = Not at all to 5 = Extremely) | 5.0 | ||
How useful in managing your diabetes was the device? (1 = Not at all to 5 = Extremely) | 4.8 | ||
How much did you trust the device? (1 = Not at all to 5 = Extremely) | 5.0 | ||
I had greater peace of mind while wearing the device. (1 = Strongly disagree to 5 = Strongly agree) | 4.8 | ||
It will be hard to give up the device once the study is over. (1 = Strongly disagree to 5 = Strongly agree) | 5.0 |