Skip to main content
. 2018 Apr 4;23(6):550–559. doi: 10.1177/2472630318762384

Table 2.

Total and Relative Area Numbers Generated during the Validation Process Using Both Protocol Versions, Including the Average Difference of Each Method from the Manual Method.

Total and Relative Areas Measured during Validation
Total Area (microV*second)
Relative Area (%)
QC
VV
QC
VV
N-Glycan Andrew 1 (N = 52) Andrew 2 (N = 40) Manual (N = 23) Andrew 1 (N = 48) Manual (N = 6) Andrew 1 (N = 52) Andrew 2 (N = 40) Manual (N = 23) Andrew 1 (N = 48) Manual (N = 6)
A2 159,358.77 150,642.29 231,777.00 181,610.26 243,364.58 1.12 1.26 1.58 1.67 1.71
FA2 6,349,848.92 5,590,096.71 6,529,769.00 4,846,368.22 6,486,709.83 44.60 44.65 44.62 44.80 45.68
FA2G1a 2,921,047.82 2,607,333.65 2,998,212.28 2,231,314.23 2,878,339.67 20.49 20.66 20.49 20.64 20.27
FA2G1b 3,340,285.09 2,964,714.69 3,387,556.09 2,455,016.26 3,310,405.25 23.42 23.43 23.15 22.70 23.31
FA2G2 1,293,861.37 1,149,093.61 1,339,268.13 996,007.11 1,202,646.08 9.06 9.06 9.15 9.21 8.47
FA2G2Sg1/FA2G2Ga2 188,310.49 123,978.16 147,138.54 105,859.73 78,596.92 1.32 0.95 1.01 0.98 0.55
Average difference from manual 11.6% 17.5% 25.1% 10.5% 4.9% 15.7%