. 2018 Aug 2;84(Suppl 1):30–45. doi: 10.1038/s41390-018-0080-2

Table 1.

Optimal setting for tissue Doppler deformation measurement

Term infants Preterm infants
Probe 5s or 7s 10s or 12s
Sector width Narrow Narrow
Sector depth Shallow Shallow
Velocity scale (cm/s) (avoid aliasing) −16 to 16 −16 to 16
Transducer frequency (MHz) 2.5–3.0 8.0
Pulse repetition frequency (kHz) 1.0 2.0
Velocity scale (cm/s) 16 16
Frame rate (FPS) >200 >200
Region of interest length (mm) 1 1
Region of interest width (mm) 2 or 3 2 or 3
Strain length (mm) 10–20 6
Linear drift compensation On On
Gaussian smoothing On On

These settings apply for the General Electric Vivid scanners and Echo Pac Software (GE Medical, Milwaukee, USA). These recommendations are a guide only and setting may differ depending on new information emerging