Table 1.

Intra-assay and inter-assay accuracy and precision tests for AS and DHA in human plasma.

Theo. Conc . Found (n = 6) Inaccuracy (%) %CV
AS DHA AS DHA AS DHA AS DHA
Intra-assay
1. LLOQ 3.20 5.33 3.02 5.53 -5.73 3.66 17.6 7.15
2. 3 x LLOQ 10.0 16.0 10.7 16.3 7.12 1.71 2.76 5.29
3. Medium 500 800 533 796 6.62 -0.55 4.00 5.28
4. 0.8 x ULOQ 2,500 4,000 2,481 3,810 -0.77 -4.74 1.68 3.11
5. ULOQ 3,000 5,000 2,782 4,637 -7.28 -7.27 3.78 4.10
Inter-assay
1. LLOQ 3.20 5.33 3.28 5.65 2.36 5.93 5.44 10.2
2. 3 x LLOQ 10.0 16.0 9.93 16.7 -0.69 4.25 8.88 5.19
3. Medium 500 800 496 785 -0.76 -1.82 7.98 5.87
4. 0.8 x ULOQ 2,500 4,000 2,445 3,872 -2.21 -3.20 5.48 4.89
5. ULOQ 3,000 5,000 2,912 4,966 -2.93 -0.68 5.03 5.24

Values are average of six replicates for intra-assay test and triplicates per day for six different days of inter-assay test.