Skip to main content
. 2018 Nov 6;18(11):3805. doi: 10.3390/s18113805

Table 3.

Deviation percentage and relative uncertainty of smartwatch 1 as compared to the reference vibrometer.

Frequency (Hz)
Amplitude 6.3 8 10 12.5 16 20 25 31.5 40 50 63 80 100
10 m/s2 ε (%) −1.14 1.28 0.63 0.00 0.81 −0.70 −1.15 −1.56 −1.31 −2.81 −4.83 −9.32 −20.64
Urel ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.1
5 m/s2 ε (%) 1.10 −0.37 −0.20 −0.29 −0.30 −0.65 −1.28 −1.65 −1.12 −2.45 −4.54 −9.13 −20.68
Urel ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.2
1 m/s2 ε (%) 1.41 0.49 0.25 0.35 −0.45 −0.63 −1.54 −2.51 −1.61 −3.03 −4.84 −9.44 −20.82
Urel ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.8

Values in red: deviation values greater than 6% with respect to the reference.