Skip to main content
. 2018 Nov 6;18(11):3805. doi: 10.3390/s18113805

Table 5.

Deviation percentage and relative uncertainty of smartwatch 3 as compared to the reference vibrometer.

Frequency (Hz)
Amplitude 6.3 8 10 12.5 16 20 25
10 m/s2 ε (%) −0.88 0.71 −0.17 0.03 −0.12 −0.29 11.90
Urel ±3.1 ±3.1 ±3.1 ±3.2 ±3.3 ±3.4 ±3.6
5 m/s2 ε (%) 0.74 −0.11 0.06 0.05 −0.04 −1.13 11.68
Urel ±3.2 ±3.1 ±3.1 ±3.2 ±3.3 ±3.4 ±3.6
1 m/s2 ε (%) −0.42 −1.50 −1.53 −1.84 −1.43 −1.41 7.41
Urel ±3.1 ±3.1 ±3.2 ±3.2 ±3.3 ±3.4 ±3.7

Values in red: deviation values greater than 6% with respect to the reference.