Skip to main content
. Author manuscript; available in PMC: 2019 Oct 1.
Published in final edited form as: IEEE Trans Biomed Circuits Syst. 2018 Sep 20;12(5):1100–1111. doi: 10.1109/TBCAS.2018.2871470

TABLE I.

Detail Summary at the Optimal Points in Fig. 5

fopt (MHz) ηfocus,TX (%) Ra (cm) w3dB (mm) [calc./sim.]
Depth (cm) 6 10 6 10 6 10 6 10
a = 1.0 cm FP 2.16 0.82* 0.65 0.16* 10.1 × 2.13/2.24 9.33/9.48*
GM 1.69 1.08 1.54 0.22 6.40 10.7 2.72/2.76 7.08/7.10
a = 2.0 cm FP 1.51 1.22 4.32 0.50 6.64 13.2 1.52/1.60 3.14/3.28
GM 1.45 1.04 4.72 0.81 6.31 10.2 1.58/1.65 3.68/3.76
a = 3.0 cm FP 1.26 1.02 7.32 1.43 6.80 11.1 1.21/1.33 2.50/2.64
GM 1.22 0.98 7.51 1.62 6.70 10.5 1.25/1.40 2.60/2.72
*

Plane transducer, off-focus operation.