TABLE III.

US Link Efficiency for a Load power of 3 mW at the Resonance of RXs Using the with a Circular Aperture size of 2 cm

fsc (MHz) ηimplant (%) ηfocus,TX (%) [FP/GM] ηUS (%) [FP/GM] w3dB (mm) calc.
Width (mm) 6 cm 10 cm 6 cm 10 cm 6 cm 10 cm
0.3 1.39 105 0.44
0.49
0.046
0.069
0.46
0.51
0.048
0.073
1.65 2.75
0.4 1.34 161 0.76
0.84
0.08
0.13
1.22
1.35
0.13
0.20
1.71 2.85
0.5 1.20 150 1.10
1.24
0.13
0.20
1.65
1.86
0.19
0.30
1.91 3.19
0.6 1.14 128 1.50
1.71
0.18
0.29
1.93
2.19
0.23
0.38
2.01 3.36
0.7 1.04 96 1.88
2.16
0.24
0.40
1.80
2.07
0.23
0.39
2.21 3.68
0.8 1.01 62 2.37
2.74
0.30
0.52
1.47
1.70
0.19
0.32
2.27 3.79
0.9 0.92 51 2.72
3.18
0.36
0.65
1.39
1.62
0.19
0.33
2.49 4.16
1.0 0.86 43 3.08
3.67
0.42
0.78
1.32
1.58
0.18
0.34
2.67 4.45
*

ηUS are calculated values based on ηimplant and ηfocus,TX.