Skip to main content
. 2018 Nov 24;13:583–596. doi: 10.1515/med-2018-0087

Table 5.

Gives percentage of bispectral index (BIS), A-line ARX Index (AAI), response entropy (RE) and state entropy (SE) values for each clinical state of the patient (state clinic) (e.g. preanesthesia awake (awake pre), postanesthesia awake (awake post), eventful anesthesia (movers), uneventful anesthesia (non-movers)) and the state as defined by the monitor (state monitor) (e.g. awake (awake), inadequate anesthesia (inadequate), optimal anesthesia (optimal), deep anesthesia (deep)).

state clinic state monitor BIS SE RE AAI
(%) (%) (%) (%)
awake pre awake 90.94 12.01 97.41 83.92
(nT = 541) inadequate 9.06 86.51 2.03 16.08
optimal 0.00 1.11 0.37 0.00
deep 0.00 0.37 0.18 0.00
anesthesia movers awake 0.97 1.58 6.62 13.99
(nT = 1,965) inadequate 9.57 11.86 16.95 40.51
optimal 40.51 43.16 38.07 29.41
deep 48.96 43.41 38.37 16.08
anesthesia non-movers awake 0.00 0.00 3.07 0.31
(nT = 12,686) inadequate 1.70 4.97 4.29 20.05
optimal 21.13 38.55 45.29 43.00
deep 77.17 56.48 47.36 36.65
awake post awake 8.11 7.66 70.72 71.17
(nT = 222) inadequate 90.54 83.78 25.68 25.68
optimal 1.35 8.56 3.60 2.70
deep 0.00 0.00 0.00 0.45