Skip to main content
. 2016 Nov 2;21(11):1464. doi: 10.3390/molecules21111464

Table 1.

Table represents the quantification data of trastuzumab signature peptide (IYPTNGYTR) in human serum batch; calibration curve (CC) extends from 5 µg/mL to 500 µg/mL; their corresponding accuracy percentage and R2 = 0.99 was observed.

Sample Name (CC) Analyte Peak Area Analyte Retention Time (RT) (min) Internal Standard (IS) Peak Area IS RT (min) Area Ratio Analyte Conc (µg/mL) Calculated Conc (µg/mL) Accuracy (%)
STD_8 2201571 4.476 1269609 4.471 1.734 501.6 526.911 105
STD_7 1764292 4.563 1184001 4.557 1.49 451.44 452.73 100.3
STD_6 1523150 4.528 1249879 4.523 1.219 375.598 370.177 98.6
STD_5 1075654 4.46 1216198 4.456 0.884 251.651 268.549 106.7
STD_4 289982 4.537 1226623 4.534 0.236 75.495 71.487 94.7
STD_3 101021 4.534 1244545 4.531 0.081 25.668 24.281 94.6
STD_2 44585 4.514 1299658 4.511 0.034 10.267 10.029 97.7
STD_1 23549 4.501 1264843 4.5 0.019 5.134 5.259 102.4