Table 2.
Raw CT numbers obtained from all 22 scanner/protocols for the 3 reference foams, air, water, and the lung foam (backing foam), along with the kVp and dose settings. A shifted CT number = (raw CT number)/1000 + 1 is used for subsequent analysis
Scanner | Foam 1 | Foam 2 | Foam 3 | Air | Water | Lung | kVp (kV) | Dose (mGy) |
---|---|---|---|---|---|---|---|---|
1 | −940.53 | −823.05 | −698.58 | −1001.40 | 1.91 | −855.72 | 100 | 1.5 |
−940.27 | −822.26 | −697.56 | −1001.53 | 0.84 | −855.69 | 120 | 1.5 | |
−940.16 | −822.86 | −698.28 | −1001.61 | 2.12 | −855.14 | 100 | 3.0 | |
−939.83 | −822.42 | −697.09 | −1001.85 | 1.30 | −855.29 | 120 | 3.0 | |
−940.11 | −823.04 | −698.18 | −1001.96 | 2.55 | −855.06 | 100 | 6.0 | |
−940.14 | −822.32 | −697.23 | −1002.13 | 1.48 | −855.24 | 120 | 6.0 | |
2 | −942.65 | −822.21 | −698.42 | −1002.70 | 3.88 | −859.86 | 120 | 3.0 |
3 | −936.48 | −820.92 | −696.00 | −995.38 | −0.91 | −852.24 | 100 | 1.5 |
−935.52 | −819.08 | −695.17 | −994.04 | −1.14 | −851.71 | 120 | 1.5 | |
−935.28 | −818.18 | −694.68 | −993.92 | −0.76 | −851.70 | 140 | 1.5 | |
−936.78 | −820.97 | −695.85 | −997.35 | −1.15 | −852.43 | 100 | 3.0 | |
−936.02 | −819.73 | −695.00 | −997.21 | −1.29 | −852.37 | 120 | 3.0 | |
−935.77 | −818.97 | −694.45 | −996.54 | −0.92 | −852.01 | 140 | 3.0 | |
−936.50 | −820.76 | −695.67 | −998.37 | −1.07 | −852.34 | 100 | 6.0 | |
−935.94 | −819.40 | −694.74 | −997.79 | −1.45 | −852.25 | 120 | 6.0 | |
−935.40 | −818.64 | −694.29 | −997.19 | −1.11 | −851.90 | 140 | 6.0 | |
4 | −940.65 | −826.08 | −699.66 | −1002.31 | 0.44 | −858.55 | 100 | 3.0 |
−940.37 | −824.64 | −699.18 | −1002.44 | 0.16 | −858.03 | 120 | 3.0 | |
−940.38 | −823.94 | −698.84 | −1002.22 | −0.08 | −857.73 | 135 | 3.0 | |
−940.59 | −825.90 | −699.83 | −1002.23 | 0.25 | −858.28 | 100 | 6.0 | |
−940.55 | −824.57 | −699.37 | −1002.36 | 0.01 | −857.88 | 120 | 6.0 | |
−940.35 | −823.86 | −699.04 | −1002.28 | −0.22 | −857.55 | 135 | 6.0 |
Raw CT number (HU) reported for each scanner‐protocol.