Table 4.
CT numbers calibrated and mapped to 80 keV from all vendor‐protocols
Scanner | Foam 1 | Foam 2 | Foam 3 | Air | Water | Lung | kVp (kV) | Dose (mGy) |
---|---|---|---|---|---|---|---|---|
1 | −939.26 | −822.01 | −697.80 | −1000.00 | 0.00 | −854.62 | 100 | 1.5 |
−939.12 | −821.85 | −697.93 | −1000.00 | 0.00 | −855.07 | 120 | 1.5 | |
−938.80 | −821.99 | −697.91 | −1000.00 | 0.00 | −854.13 | 100 | 3.0 | |
−938.51 | −822.13 | −697.89 | −1000.00 | 0.00 | −854.70 | 120 | 3.0 | |
−938.49 | −822.08 | −697.92 | −1000.00 | 0.00 | −853.91 | 100 | 6.0 | |
−938.60 | −821.90 | −698.01 | −1000.00 | 0.00 | −854.51 | 120 | 6.0 | |
2 | −940.44 | −820.96 | −698.19 | −1000.00 | 0.00 | −858.33 | 120 | 3.0 |
3 | −940.34 | −823.32 | −696.82 | −1000.00 | 0.00 | −855.05 | 100 | 1.5 |
−940.68 | −822.69 | −697.12 | −1000.00 | 0.00 | −855.76 | 120 | 1.5 | |
−940.68 | −822.27 | −697.37 | −1000.00 | 0.00 | −856.16 | 140 | 1.5 | |
−939.19 | −822.91 | −697.29 | −1000.00 | 0.00 | −854.51 | 100 | 3.0 | |
−938.78 | −822.44 | −697.65 | −1000.00 | 0.00 | −855.10 | 120 | 3.0 | |
−939.18 | −822.29 | −697.66 | −1000.00 | 0.00 | −855.36 | 140 | 3.0 | |
−938.21 | −822.61 | −697.67 | −1000.00 | 0.00 | −854.15 | 100 | 6.0 | |
−938.35 | −822.17 | −697.90 | −1000.00 | 0.00 | −854.92 | 120 | 6.0 | |
−938.39 | −821.98 | −698.00 | −1000.00 | 0.00 | −855.14 | 140 | 6.0 | |
4 | −938.28 | −823.61 | −697.06 | −1000.00 | 0.00 | −856.11 | 100 | 3.0 |
−938.11 | −822.72 | −697.62 | −1000.00 | 0.00 | −856.01 | 120 | 3.0 | |
−938.40 | −822.39 | −697.76 | −1000.00 | 0.00 | −856.06 | 135 | 3.0 | |
−938.27 | −823.42 | −697.18 | −1000.00 | 0.00 | −855.85 | 100 | 6.0 | |
−938.32 | −822.59 | −697.65 | −1000.00 | 0.00 | −855.83 | 120 | 6.0 | |
−938.30 | −822.23 | −697.87 | −1000.00 | 0.00 | −855.80 | 135 | 6.0 |
CT numbers80keV = (H 80 kev − 1) × 1000.
CT number (HU) mapped to 80 keV for each scanner‐protocol.