Table 1.

Detailed description of the subtests included in the 5 tests performed to both measure the accuracy of QCPR cam-app 2.0's ability to detect the compression rate under various conditions and to evaluate the CPR summary parameters calculated after an ended session.

Subtest name Compression rate (cpm) Duration (s) Camera position Lighting Measures
Smartphone position test (n=7)
RateP126 Normal 60, 100, 120, 150 60 x 4 Pos.1 480 lux μE¯, μP
D1R110P126 Disturb. person 110 120 Pos.1 480 lux μE¯, μP
30:2P1 30:2 110 90 Pos.1 480 lux μE¯, μP
LightP1 Dimmed light 110 60 Pos.1 7 lux μE¯, μP
R100P2 Small part of image frame (position Change) 100 60 Pos.2 480 lux μE¯, μP
R150P2 Small part of image frame (position Change) 150 60 Pos.2 480 lux μE¯, μP
30:2P3 30:2 (position Change) 110 90 Pos.3 480 lux μE¯, μP
R100P3 Normal (position Change) 100 60 Pos.3 480 lux μE¯, μP

Outdoor test (n=3)
OCBR110P1 Cloudy with noisy (threes) background 110 60 Pos.1 Cloudy weather μE¯, μP
OCR110P1 Cloudy with no background 110 60 Pos.1 Cloudy weather μE¯, μP
OSBR110P1 Sunny with noisy (threes) background 110 60 Pos.1 Sunny weather μE¯, μP
OSR110P1 Sunny with no background 110 60 Pos.1 Sunny weather μE¯, μP

Disturbance test (n=1)
D2R110P1 Disturbing person 110 180 Pos.1 Normal indoor μE¯, μP

Random movement test (n=3)
Ran.MovP1 Random movements 150 Pos.1 Normal indoor μP

CPR summary report test (n=5)
CPRsrR110P1 Compressions with pauses 110 580 Pos.1 Normal indoor μREpar

R = rate; P = position; D = disturbance; O = outdoor; B = noisy background; C = cloudy; S = sunny; CPRsr = CPR summary report.