Table 1.

α-SMA ELISA Technical Validation Data

Technical Validation Test Result
IC50 3.43 ng/ml
Detection range 0.44-25.5 ng/ml
Intra-assay variation 8.0%
Inter-assay variation 13.0%
Dilution recovery of human serum 99%
Dilution recovery of human urine 82%
Dilution recovery of rat serum 84%
Dilution recovery of rat urine 90%
Analyte recovery 24 h, 4°C/20°C 102%/97%
Hemoglobin recovery 86%
Lipemia recovery 99%
Analyte recovery, 4 freeze/thaw cycles 97%
Salt recovery, pH 7.0/pH 8.0 99%

Percentages are reported as mean.

Average recovery after salt interference.