Skip to main content
. 2018 May;19(5):341–355. doi: 10.14744/AnatolJCardiol.2018.76736

Table 1.

Interventions to stenotic outflow grafts: a summary

Reference Pre-Interventional flow and power (L/W) Pre-Interventional rate (RPM) Post-Interventional flow and power (L/W) Post-Interventional rate (RPM) Intervention Cerebral protection Device On/Off Stenosis severity
2 2.2/4.9 3600 5/6 3200 Covered stent implantation None Off Unknown
3 2.5/Unknown 2900 4/Unknown 2350 Covered stent implantation None On 90%
4 1/2 2700 6/4.9 2700 Bare-metal stent implantation None On Nearly occluded
4 3/2.5 3000 6/5.9 3000 Bare-metal stent implantation None On Nearly occluded
4 2/1 2500 4/3 2500 Bare- metal stent implantation Balloon blockade On occluded
5 8/Unknown Unknown Unknown Unknown Covered stent implantation Balloon blockade + embolic filter Off Unknown
* 0 2500 3.7/2.9 2500 Balloon dilatation only Manual carotid occlusion Off Total occlusion

L - Liters; W - Watts; RPM - Revolutions per minute

*

Our patient