Type of institution |
Private |
Public |
Small |
Medium |
Large |
Total amount of X-ray rooms |
Total amount of equipment in the institution |
Fixed conventional X ray |
Portable conventional X ray |
Mammography |
Tomography |
Fluoroscopy |
Processer |
Number of professionals in the institution |
Medical radiologists |
Medical residents |
Technicians with radiology course |
Technicians without radiology course |
Amount of technicians with experience in radiology? |
How many with experience <2 y |
How many with experience of 2–5 y |
How many with experience of 5–10 y |
How many with experience >10 y |
Does the facility have a medical physicist? |
Is there a quality control programme implemented? |
Is it performed by the medical physicist? |
Is it performed by the technician? |
Is it performed by an external engineer? |
Is IAEA protocol used? |
Is ACR protocol used? |
Is Spanish protocol used? |
Other? |
Assessment of radiological equipment |
Equipment brand |
Equipment model |
Tube brand |
Tube model |
Year of manufacture |
Installation date |
Range of kV |
Range of mA s |
Is the equipment analogical? |
Is the equipment digital? |
Automatic exposure control? |
Focal spot marked on housing? |
Is the total filtration value indicated? |
Is there an audible or luminous indication on the command console to reveal radiation emission? |
Is it possible to see and maintain contact with the patient? |
Is the luminous indication of the radiation field functioning? |
Equipment maintenance |
Is there a preventive maintenance programme? |
Is it internal? |
Supplied by the manufacturer or representative? |
Supplied by an external company? |
Is there a contract? |
Maintenance frequency: biannual? |
Maintenance frequency: annually? |
Maintenance frequency: every 2 y? |
Other? |
Environmental conditions of the room |
Is there air conditioning in the mammography equipment room? |
Is the air conditioning functioning? |
Does the room have reinforcement? |
Are the doors reinforced? |
Is there a luminous indicator on the door? |
Is there a radioprotection sign on the door? |
Protection measures for the operator and the patient |
Is there a fixed barrier or window lead shielding? |
What material is the barrier made of? |
Are there protection aprons in the room? |
How many? |
Do technicians use a personal dosemeter? |
Is it a film dosemeter? |
Is it TLD? |
Other dosemeters? |
Assessment of processing systems |
Do you use an X-ray film to obtain image? |
Do you use CR to obtain image? |
Do you use DR to obtain image? |
In the case of Digital Radiology: |
Model of CR cassettes? |
Brand of CR cassettes? |
Model of DR detector? |
Brand of DR detector? |
Can the images be extracted? |
In ‘for processing’ format? |
In ‘for processing’ format? |
Brand of the workstation monitor? |
Model of the workstation monitor? |
Is the workstation monitor CRT type? |
Is the workstation monitor LCD type? |
Matrix size of workstation monitor? |
Does the software of the monitor belong to the workstation? |
In the case of radiographic film |
Brand? |
Model? |
Type of screen? |
Is the process of the film manual? |
Is the process of the film automatic? |
Brand of processer? |
Model of processer? |
Developer temperature? |
Total processing time? |
Is the variation between the value of measured time and the one indicated on the equipment acceptable? (Tolerance: <±3 % with respect to the value indicated by the manufacturer) |
If you have a thermometer measure the temperature of the developer |
Is the variation between the measured value and the one indicated on the equipment acceptable? (Tolerance: ±0.5 with respect to the value indicated by the manufacturer) |
Is there evidence of stains or scratches in the films after the process? |
Darkroom inspection |
Is there a ventilation system? |
Is it functioning? |
Is the floor non-slip, waterproof and anticorrosive? |
Is the room clean? |
Can you perceive a strong odour of liquid developer? |
Is there an extractor fan? |
Is the white light switch located in places where it cannot be activated involuntarily? |
Is the safety light at a distance of more than 120 cm from the surface? |
Is the power of the safety light lower or equal to 15 W? |
Are there signs of infiltration or dampness on the walls? |
Are the boxes of film in vertical position and organised? |
Are there entries of light in the darkroom? |
Are the films organised by their expiry date? |
Amount of negatoscopes |
Can you observe stains at a first glance? |
Are there luminosity differences in the negatoscope? |
Is there a magnifying glass? |
Can you vary the intensity of the negatoscope? |
Can you disguise the image of the X ray? |
Can you vary the intensity of the observation room? |
If you have a photometer, measure the luminance of the negatoscope |
Is the measured value acceptable? (tolerance: luminance: 1500 cd m−2) |
Uniformity of the negatoscope |
Is the measured value acceptable? (Tolerance: Uniformity: <15 %) |
Illumination of the viewing room |
Is the measured value acceptable? (tolerance: illumination: <100 lux) |