Skip to main content
. 2018 Nov 6;32(23):1999–2023. doi: 10.1002/rcm.8272

Table 2.

Mean concentrations, coefficient of variation (CV), and mean accuracy of measured steroid hormones in walrus bone tissue spiked with each hormone's respective concentrations along a standard calibration curve (i.e., “Concentration (ng)”). Respective concentrations for each hormone have n = 5 replicates. Accuracy was determined following Zhang et al45

Hormone Concentration (ng) Mean concentration (ng) CV (%) Mean accuracy (%)
Progesterone 3200 3206.04 9.33 100.19
1600 1535.04 0.98 95.94
800 869.73 6.50 108.72
400 445.19 6.69 111.30
200 199.90 6.11 99.95
100 102.58 5.70 102.58
50 50.13 7.67 100.25
0.5 0.53 3.39 105.54
Testosterone 500 499.76 1.84 99.95
250 251.30 4.17 100.52
125 124.73 2.19 99.78
75 71.92 1.61 95.92
0.5 0.54 3.15 107.07
Cortisol 500 498.09 4.03 99.62
250 255.25 2.99 102.10
125 125.19 1.10 100.15
75 69.65 3.14 92.87
0.5 0.54 8.30 107.50
Estradiol 200 200.90 2.64 100.45
100 100.22 3.54 100.22
50 48.99 7.96 97.98
25 25.24 3.64 100.97
5 5.40 8.56 107.99