. 2018 Oct 24;37:382–391. doi: 10.1016/j.ebiom.2018.10.012

Table 4.

Specificity results with the negative samples (N = 200).

HST/HT EXACTO PRO TEST HIV EZ-TRUST HIV 1 & 2 rapid screen Test Genie fast HIV1/2 HIVTOP
INSTI STAT-VIEW HIV 1/2 VIKIA HIV 1/2 BioTechMed HIV1/2 Rapid-4 a
HIV combo
T1 T2 T1 T2 Ag Ac
Nbr of positive samples 3 0 0 2 0 0 0 1 0 0 1 0
Specificity 98,50% 100% 100% 99% 100% 100% 100% 99,50% 100% 100% 99,50% 100%
95% CI (%) 95·3–99·6 97·6–100 97·6–100 96·0–99·8 97·6–100 97·6–100 97·6–100 96·8–99·9 85·9–100 85·9–100 96·8–99·9 97·6–100

Specificity was evaluated on 200 HIV-negative whole blood samples with seven 3G RSTs/STs (left part of the table) and two 4G RSTs/STs (right part of the table). The table reports the false-positive results. For the HIV Combo (Alere) test, the detection of p24Ag is performed with a specific line, distinct from the antibody line. No antigen-specific line was present in the BioTechMed HIV1/2 Rapid-4 test. HIVTOP and BioTechMed HIV1/2 Rapid-4 are designed to discriminate HIV-1 from HIV-2 infections; results for both HIV-1 and HIV-2 are reported.

a

N = 30 samples.