Table 2.
Device-related complications
N (%) | TV-ICD (n = 46) | S-ICD (n = 35) |
---|---|---|
Endocarditis | 1 (2%) | – |
Lead displacement | 5 (11%) | – |
Lead dysfunction | 5 (11%) | – |
Local device infection | – | 3 (9%) |
Failed conversion test | – | 1 (3%) |
Inadequate sensing | 1 (2%) | 1 (3%) |
Pocket erosion | – | 1 (3%) |
Fistula between left mammary artery and subclavian vein | 1 (2%) | – |