Table 1.
Technical specification of magnetic resonance imaging |
---|
For MRI-conditional devices follow manufacturer’s instructions for scan |
For non-MRI-conditional devices: |
– 1.5 T MRI should be preferred |
– Gradient slew rate should not exceed 200 T/m/s |
– Dorsal patient position |
– Imaging landmark near the device (thorax) should be avoided |
– Local transmit coils should be avoided |
– SAR and scan time should be limited |
Emergency equipment/external defibrillator as well as a device programmer should be present during the MRI |
Continuous patient monitoring (electrocardiography/pulse oximetry) during the MRI should be performed |
MRI magnetic resonance imaging, SAR specific absorption rate