. 2018 Dec 6;9:1721. doi: 10.3389/fphys.2018.01721

Table 5.

Light-adapted 3.0 flicker ERG (28.3 Hz) measurements recorded from patients with cognitive impairment in comparison with the normative data of the RETeval system.

Light Adapted Test (flicker cone–3.0 cd.s/m2, 30 cd/m2, 28.3 Hz) ERG reference data n = 244 median [90% CI] Age: [76 – 86] Cognitive Impairment Cases n = 20 median [90% CI] Age: [69 – 90] Cognitively Healthy Cases n = 19 median [90% CI] Age: [60 – 88] p-value
Amplitude(μV) 2.5% limit
19.6 [18–22.1]
97.5% limit
39.4 [35.5–43.1]
12.6 [10.9 – 14.4] 37.4 [36.6 – 38.2] <0.001∗∗
Implicit Time(ms) 2.5% limit
25.6 [25.1–26.3]
97.5% limit
29.6 [29.2–29.9]
31.2 [29.6–32.8] 29.6 [29.4 –29.8] <0.001∗∗

Amplitude (μV) and implicit time (ms) are denoted along with the medians and 90% confidence intervals of the 2.5 and 97.5% reference limits. The P-values comparing cognitive impairment vs. cognitively healthy cases were calculated by independent sample t test. ∗∗Significant (p < 0.01).