Table 1. Summary of optimization objectives for CyberKnife, Elekta, and Varian linear accelerator plans.
Structure | Criteria | Dose Objective (Gy) |
PTV | Min Dose | 25.0 |
PTV | Min DVH D95 | 26.5 |
PTV | Max Dose | 32.0 |
RT Coronary Artery | Max Dose | 19.0 |
Ramus Coronary Artery | Max Dose | 19.0 |
LT Anterior Descending CA | Max Dose | 19.0 |
Mitral Valve | Max Dose | 19.0 |
Pericardium | Max Dose | 19.0 |
LT Main CA | Max Dose | 19.0 |
Circumflex | Max Dose | 19.0 |
LT Bronchus | Max Dose | 20.0 |
Defibrillator | Max Dose | 2.0 |