Published in final edited form as: Phys Med Biol. 2018 Mar 7;63(5):055015. doi: 10.1088/1361-6560/aaaf01

Table 4.

The acoustic power measured in Watts for each transducer and drive power setting. Uncertainties indicate the standard deviation from repeat measurements (4 repeats for ‘low’, 3 repeats for ‘moderate’ and ‘high’ drive settings).

Nominal drive setting Acoustic Power (W) for each Transducer
1 2 3 4a 4b 5 6 7
‘low’ – 50mV pk-pk 1.4±0.07 0.91±0.09 1.14±0.18 1.18±0.08 1.86±0.33 1.31±0.05 1.20±0.13 1.36±0.11
‘moderate’ – 150mV pk-pk 11.86±0.21 8.09±0.51 9.15±0.88 10.59±0.42 15.73±1.23 11.60±0.50 10.12±0.60 14.43±0.38
‘high’ 59.43±1.04 41.47±0.15 20.42±1.03 52.38±0.72 68.28±0.36 24.67±0.12 56.78±0.48 69.97±0.80