Published in final edited form as: J Magn Reson Imaging. 2018 Aug 24;49(1):291–303. doi: 10.1002/jmri.26203

Table 3.

Repeatability analysis for mean placental R2* measured in normal subjects using free-breathing radial MRI at 3 T. The within-technique mean difference (MDwithin), absolute within-technique mean difference (MDabs), within-technique standard deviation (SDwithin), and coefficient of repeatability (CR) for placental R2* measured by two back-to-back scans were calculated for all normal subjects (N=30) and again for the subjects separated into anterior (N = 15) versus posterior (N = 15) placenta implantation positions. The analysis was performed for each gestational age (GA) range.

GA (Weeks) MDwithin (s−1) MDabs (s−1) SDwithin (s−1) CR (s−1)
All Subjects
(N = 30)
14-18 0.32 0.84 1.05 2.92
19-23 0.98 1.73 2.97 8.241
Anterior
(N = 15)
14-18 0.48 1.05 1.21 3.37
19-23 0.24 1.23 1.66 4.60
Posterior
(N = 15)
14-18 0.15 0.63 0.87 2.42
19-23 1.73 2.23 3.79 10.501
1

Includes two severe outliers during the second scan (see Discussion). By removing these outliers, the CR for 19-23 weeks was 4.20 s−1 (all subjects) and 3.85 s−1 (posterior placenta).