Skip to main content
. 2018 Dec 17;6:24–32. doi: 10.1016/j.ejro.2018.12.001

Table 3.

Volume CT dose indices (CTDIvol) and tube voltages at different phantom’s vertical positions in routine chest CT protocols with contrast administration using ATVS. The tube voltages changed compared to the reference table height are bolded.

Patient vertical position (cm) AP scout (Siemens)
AP scout (GE)
PA scout (Siemens)
PA scout (GE)
LAT scout (Siemens)
LAT scout (GE)
CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp)
+6 4.22 120 3.15 100 3.75 120 1.70 100 2.00 80 2.54 100
+3 3.89 120 2.60 100 4.20 120 1.94 100 1.95 80 2.53 100
0 2.61 100 2.17 100 4.51 120 2.28 100 1.91 100 2.51 100
−3 2.37 100 1.87 100 4.87 120 2.86 100 1.94 100 2.47 100
−6 2.20 100 1.63 100 5.21 120 3.43 100 1.96 80 2.41 100

CTDI = Computed tomography dose index, AP = Anterior-to-posterior, PA = Posterior-to-anterior, LAT = Lateral.