Volume CT dose indices (CTDIvol) and tube voltages at different phantom’s vertical positions in routine chest CT protocols without contrast administration using a fixed 120 kVp tube voltage.

Patient vertical position (cm) AP scout (Siemens)
AP scout (GE)
PA scout (Siemens)
PA scout (GE)
LAT scout (Siemens)
LAT scout (GE)
CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp) CTDI (mGy) Tube voltage (kVp)
+6 4.22 120 3.70 120 3.83 120 1.99 120 2.62 120 2.96 120
+3 3.89 120 3.06 120 4.22 120 2.29 120 2.56 120 2.98 120
0 3.58 120 2.57 120 4.62 120 2.68 120 2.48 120 3.00 120
−3 3.27 120 2.21 120 4.79 120 3.33 120 2.51 120 2.91 120
−6 2.98 120 1.93 120 5.24 120 4.03 120 2.53 120 2.82 120

CTDI = Computed tomography dose index, AP = Anterior-to-posterior, PA = Posterior-to-anterior, LAT = Lateral.