Table 3.
Tecnical validation step | SPARC-M performance |
---|---|
Detection range (LLOD-ULOD) | 2.7–300.7 ng/mL |
Intra-assay variation | 6% |
Inter-assay variation | 10% |
Dilution of serum samples | 1:4 |
Dilution recovery (1:4 pre-dilution) | 96% (77–102%) |
Freeze/thaw recovery (4 cycles) | 92% (86–103%) |
Analyte stability up to 48 h, 4°C and 4 h, 20°C | 88% (84–96%) |
Interference Lipids, low/high | 96%/97% |
Interference Biotin, low/high | 96%/98% |
Interference Hemoglobin, low/high | 96%/80% |
Percentages are reported as mean with range shown in brackets