Skip to main content
. 2018 Aug 1;19(10):904–912. doi: 10.1080/15384047.2018.1480887

Table 3.

Technical validation data of the SPARC-M ELISA assay.

Tecnical validation step SPARC-M performance
Detection range (LLOD-ULOD) 2.7–300.7 ng/mL
Intra-assay variation 6%
Inter-assay variation 10%
Dilution of serum samples 1:4
Dilution recovery (1:4 pre-dilution) 96% (77–102%)
Freeze/thaw recovery (4 cycles) 92% (86–103%)
Analyte stability up to 48 h, 4°C and 4 h, 20°C 88% (84–96%)
Interference Lipids, low/high 96%/97%
Interference Biotin, low/high 96%/98%
Interference Hemoglobin, low/high 96%/80%

Percentages are reported as mean with range shown in brackets