Skip to main content
. 2018 Dec 1;18(12):4218. doi: 10.3390/s18124218

Table 2.

Comparison of the EIS response of the aptasensor in a PBS and in spiked milk samples. Recovery was calculated as: [(ΔRct/Rcto)milk/(ΔRct/Rcto)PBS] × 100%.

Concentration of AFM1, ng/L Buffer
ΔRct/Rcto
Milk
ΔRct/Rcto
Recovery, %
15 0.16 0.17 106.25%
30 0.41 0.32 78.04%
60 0.61 0.55 90.16%
90 0.65 0.63 97.00%
120 0.78 0.62 79.5%