Skip to main content
. 2018 Dec 2;18(12):4227. doi: 10.3390/s18124227

Table 4.

Accuracy evaluation based on the British Hypertension Society (BHS) standard.

CP at ± 5 mmHg CP at ± 10 mmHg CP at ± 15 mmHg Grade
Proposed model
(model 22)
SBP 50.95% 81.18% 94.77% B
DBP 64.45% 93.44% 98.76% A
PTT-based model
(model 25)
SBP 47.53% 77.28% 93.16% C
DBP 58.84% 89.62% 97.95% B