Table 2.
OCT | IVUS | |
---|---|---|
Frame rate (s) | 100 | 30 |
Pullback speed (mm/s) | 20 | 0.5–1.0 |
Lines per frame | 500 | 256 |
Axial resolution (μm) | 12–15 | 150 |
Scan diameter in contrast (mm) | 10 | 15–20 |
Tissue penetration (mm) | 1.0–2.0 | 10 |
Ease of use | ++ | ++/+++ |
Need for contrast or dextran | Yes | No |
Detection of lipid | +++ | +/++ |
Detection of fibrous cap | +++ | + |
Detection of thrombus | ++ | + |
Detection of calcium | ++ | +++ |
Dissection | +++ | ++ |
Malapposition | +++ | ++ |
Stent strut surface coverage | +++ | + |
IVUS, intravascular ultrasound; OCT, optical coherence tomography.