Skip to main content
. 2018 Aug 18;70(Suppl 3):S456–S465. doi: 10.1016/j.ihj.2018.08.012

Table 2.

Comparison: optical coherence tomography versus intravascular ultrasound.

OCT IVUS
Frame rate (s) 100 30
Pullback speed (mm/s) 20 0.5–1.0
Lines per frame 500 256
Axial resolution (μm) 12–15 150
Scan diameter in contrast (mm) 10 15–20
Tissue penetration (mm) 1.0–2.0 10
Ease of use ++ ++/+++
Need for contrast or dextran Yes No
Detection of lipid +++ +/++
Detection of fibrous cap +++ +
Detection of thrombus ++ +
Detection of calcium ++ +++
Dissection +++ ++
Malapposition +++ ++
Stent strut surface coverage +++ +

IVUS, intravascular ultrasound; OCT, optical coherence tomography.