. 2018 Dec 10;2018:3258396. doi: 10.1155/2018/3258396

Table 2.

Pressure drop characteristics of tested and control endotracheal tubes at 60 L/minute.

Type of endotracheal tube n Mean pressure drop Standard deviation Min. pressure drop Max. pressure drop t-test value vs. control p value t-test value vs cleared tube p value
All endotracheal tubes
All controls 3 10.8 0.27 n/a −8.63 <0.001
Extubated, not cleared (pre) 40 21.8 13.30 10.2 77.8 5.27 <0.001 4.80 <0.001
Extubated, cleared (post) 40 11.8 0.29 9.5 16.8 −8.63 <0.001 n/a
7.0 mm diameter
Control 1 14.5 0.14 n/a −3.69 0.035
Extubated, not cleared (pre) 4 19.2 1.88 16.4 20.4 4.94 0.016 5.20 0.014
Extubated, cleared (post) 4 15.9 0.75 15.0 16.8 −3.69 0.035 n/a
7.5 mm diameter
Control 1 11.8 0.01 n/a −3.30 0.004
Extubated, not cleared (pre) 17 24.1 9.12 12.0 46.0 4.33 <0.001 4.13 0.001
Extubated, cleared (post) 17 12.5 0.86 11.4 14.3 −3.30 0.004 n/a
8.0 mm diameter
Control 1 9.1 0.04 n/a −10.76 <0.001
Extubated, not cleared (pre) 19 23.0 17.42 10.2 77.8 3.55 0.002 3.23 0.004
Extubated, cleared (post) 19 10.3 0.48 9.5 11.3 −10.76 <0.001 n/a